Installation Instructions

Linear Encoders - Using Alignment Tool - Alignment

Page 5

Proper sensor alignment may require minor adjustments to the sensor position with
respect to the scale. This can be performed easily using the SmartPrecision Alignment
Tool as illustrated below.

The red, yellow, or green LED will light depending on sensor alignment. Slowly
move the sensor by allowing it to slide on the mounting surface until the green
LED, is illuminated. Optimal alignment will be displayed as a “Bright Green” LED.

IMPORTANT:

Confirm that the green LED blinks when passing over the index.

If not, readjust the sensor in the Y direction and repeat the above procedure.
When alignment is completed, tighten the sensor mounting screws
(0.37Nm [3.3 inch-lbs.] maximum torque).

6

5

Confirm green over the full range of
motion by sliding the scale past the sen-
sor. The “green” LED must remain on over
the entire range. If not aligned over the
entire range of motion, loosen the sensor

x

Y

Z

θ

z

To align the sensor, move
it in the Y or

θ

z direc-

tions.

SmartPrecision
Alignment Tool

Improper

Alignment LED

Red

Power/

Calibration

Power/

Calibration

Power/

Calibration

Improved

Alignment LED

Yellow

Proper

Alignment LED

Green

Optimal

Alignment LED

Bright Green

10 pin H

connector

to sensor

10 pin H

connector

to sensor

10 pin H

connector

to sensor

SmartPrecision
Alignment Tool

SSAT1500H

SmartPrecision
Alignment Tool

SSAT1500H

SmartPrecision
Alignment Tool

SSAT1500H