Measurement Computing WBK11A User Manual
Page 3

WBK11A – SSH Card
988396
WBK11A, pg. 3
Accuracy and Noise Specifications
WaveBook/516 Series
(Alone)
WaveBook/516 Series
with a WBK11A
(Note 3)
Voltage
Range
Accuracy
(Note 2)
One Year, 18-28
°C
Input Noise
LSB rms
DC-500KHz
(typical)
(
Note 4)
Accuracy
(Note 2)
One Year, 18-28
°C
Input Noise
LSB rms
DC-500KHz
(typical)
± %
reading
± %
range
± %
reading
± %
range
0 to +10V
.012%
.008%
2
.012%
.008%
2
0 to +5V (10A)
0 to +4V (516)
.012%
.009%
2
.012%
.009%
2
0 to +2V
.012%
.012%
3
.012%
.012%
3
0 to +1V
(10A only)
.012%
.018%
3
.012%
.018%
3
0 to +.5V
.018%
.033%
6
0 to +.2V
.018%
.08%
8
0 to +.1V
.018%
.16%
15
-10 to +10V
.012%
.008%
2
.012%
.008%
2
-5 to +5V
.012%
.008%
2
.012%
.008%
2
-2 to +2V
.012%
.009%
2
.012%
.009%
2
-1 to +1V
.018%
.012%
3
.018%
.012%
3
-.5 to +.5V
(10A only)
.018%
.018%
5
.018%
.018%
6
-.2 to +.2V
.018%
.033%
8
-.1 to +.1V
.018%
.08%
15
-.05 to +.05V
(10A only)
.018%
.16%
26
Notes: 1. Specifications assume differential input scan, unfiltered.
2. Accuracy specification is exclusive of noise.
3. Unipolar ranges unavailable for a WaveBook/516, /516A, or /516E
that has a WBK11A, WBK12A, or WBK13A option installed.
Unipolar mode is available with WBK10A and any option.
4.
Maximum limit is 1.3X typical.