Hardware setup, Wbk11a – specifications, Configuration – Measurement Computing WBK11A User Manual
Page 2: Installation

pg. 2, WBK11A
988396
WBK11A – SSH Card
Hardware Setup
Configuration
All WBK11A configurations are controlled by software. There are no hardware settings.
Reference Note:
For detailed WaveView information refer to the
WaveView Document Module
. A PDF version
of the document can be accessed from the data acquisition CD via the
the CD’s opening screen.
Installation
There is no user installation permitted. See notes on page 1.
WBK11A – Specifications
Name/Function: WBK11A; 8-Channel Simultaneous Sample-and-Hold Card
Number of Channels: 8
Connectors: Internal to the WaveBook/512 series, or WaveBook/516 series device;
a 36 pin-socket connector mates with a 36-pin connector
Accuracy: ±0.025% FS for WaveBook/512A applications.
For WaveBook/516 Series applications, see the table on page 3 of this WBK11A document module.
Offset: ±1 LSB max
Aperture Uncertainty: 75 ps max
Voltage Droop: 0.1 mV/ms max
Maximum Signal Voltage: ±5.00 VDC (×1)
Input Voltage Ranges:
Before a scan sequence begins, the input voltage ranges can be programmed via software.
The ranges can be expanded as follows:
Unipolar:
Unipolar applies to WBK 10A only.
0 to +10 V
0 to +5 V
0 to +2 V
0 to +1 V
0 to +0.5 V
0 to +0.2 V
0 to +0.1 V
Bipolar: Bipolar applies to
WaveBook/516, /516A, /516E, /512A, and WBK10A
-10 to +10 V
-5 to +5 V
-2 to +2 V
-1 to +1 V
-0.5 to +0.5 V
-0.2 to +0.2 V
-0.1 to +0.1 V
-.05 to + .05 V
This range applies to WBK 10A only.
Programmable Gain Amplifier Gain Ranges: ×1, 2, 5, 10, 20, 50, 100
Weight: 0.14 kg (0.3 lb)