USB-1608HS-2AO User's Guide

Functional Details

20

SYNC terminals (SYNC_IN and SYNC_OUT)

You can use the SYNC_IN connection to externally pace the A/D conversions. The SYNC_IN terminal
supports TTL-level input signals of up to 250 kHz.

Use the SYNC_OUT connection to output the clock used for A/D conversions.

One example of the use of these two pins would be to synchronize with a second USB-1608HS-2AO and
acquire synchronized data from 16 channels. You can connect the SYNC_OUT pin of one USB-1608HS-2AO
to the SYNC_IN pin of another USB-1608HS-2AO to acquire data synchronously from 16 channels.

Figure 10. Configuring for synchronous data acquisition

Trigger terminal (TRIG_IN)

The

TRIG_IN

connection is an external analog/digital trigger input.

With the analog trigger function, you can start and control acquisitions with an analog signal. The analog trigger
threshold is from -10 V to +10 V on the TRIG_IN pin. A 12-bit DAC sets the level for the threshold. The
threshold resolution in this mode is 4.88 mV.

The USB-1608HS-2AO has three trigger options that you must set.

Trigger above or trigger below
Level-sensitive or edge-sensitive
Retrigger on or retrigger off

Each trigger operation mode is explained next. In each case, a ±2 V triangle waveform is used as the TRIG_IN
input source. The high threshold is set to 1.0 V, and the low threshold signal is set to -1.0 V.

In the following analog trigger signal diagrams, the bold portion of the waveform indicates the data acquired for
the given analog trigger mode.

SYNC_IN pin

SYNC_OUT pin