beautypg.com

Analog output – Measurement Computing USB-1408FS User Manual

Page 21

background image

USB-1408FS User's Guide

Specifications

21

Note 4:

Extrapolating the long term drift accuracy specifications will provide the approximate long term drift
of the USB-1408FS intermediate input ranges.

Table 2. Accuracy, differential mode

Range

Absolute Accuracy 25 °C (±mV)

Absolute Accuracy 0 °C to 50 °C (±mV)

±20 V

10.98

49.08

±10 V

7.32

33.42

±5 V

3.66

20.76

±4 V

2.92

19.02

±2.5 V

1.83

14.97

±2 V

1.70

14.29

±1.25 V

1.21

12.18

±1 V

1.09

11.63

Table 3. Accuracy, single-ended mode

Range

Absolute Accuracy 25 °C (±mV)

Absolute Accuracy 0 °C to 50 °C (±mV)

±10 V

10.98

49.08

Table 4. Noise performance, differential mode

Range

Typical counts

Least significant bit

root mean square

(LSB

rms)

±20 V

8

1.21

±10 V

8

1.21

±5 V

9

1.36

±4 V

10

1.51

±2.5 V

12

1.81

±2 V

14

2.12

±1.25 V

18

2.72

±1 V

22

3.33

Table 5. Noise performance, single-ended mode

Range

Typical Counts

LSB

rms

±10 V

8.0

1.21

Analog output

Table 6. Analog output specifications

Parameter

Conditions

Specification

Resolution

12-bits, 1 in 4096

Output range

0 V to 4.096 V, 1 mV per LSB

Number of channels

2

Throughput (Note 5)

Software paced

250 S/s single channel typical, PC-dependent

Single channel, continuous scan

10 kS/s

Dual channel, continuous scan,
simultaneous update

5 kS/s

Power on and reset voltage

0V, ±20 mV typ (initializes to 000h code)

Output drive

Each D/A OUT

±15 mA

Slew rate

0.8 V/µs typ

Note 5:

Maximum throughput scanning from PC memory is machine dependent.