Ivoclar Vivadent VP3 easy User Manual

Page 10

background image

10

7.

Maintenance and Cleaning

For safety reasons, disconnect power plug before

maintenance and cleaning.

7.1 Monitoring and Maintenance

The time for maintenance procedures depends on the
frequency of use and the location of the pump. For that
reason the recommended times are only approximates.

This apparatus has been developed for typical use

in the dental laboratory.

If the product is used in a production facility, for industrial
applications, or in continuous operation, premature ageing
of certain spare parts have to be expected.

These spare parts are e.g.:

Sound absorber

Diaphragm

Valve reed

These spare parts are not covered by the warranty.
Please also observe the shorter service and maintenance
intervals.

7.2 Cleaning

Remove dust from the pump from time to time. Use a
vacuum cleaner with a cleaning brush for that purpose.

7.3 Notes on maintenance

The diaphragm may only be changed by an

Ivoclar Vivadent Service Center. The fuse must be
changed by qualified personnel.

Please observe the relevant rules and regulations for the
prevention of accidents as well as the safety and medical
regulations. The connection cover that is marked with the
label shown below may only be opened by qualified per-
sonnel after the power has been disconnected.

What

Check all plug-in
connections for
correct fit.

Check the hose
to make sure that
no condensation
water has
accumulated in
the system.

Part

Connections

Vacuum hose

When

Weekly

Weekly

Item

Housing

Sound absorber
(4)

Frequency

if required

if it is dirty

Cleaning material

soft, dry cloth

sound absorber
has to be
replaced

VP easy-BA-WE1.qxd 5.10.2005 11:29 Uhr Seite 10