Programat, Short instructions – Ivoclar Vivadent CS - Short Instructions User Manual
Page 2

Programat
®
CS
– Short Instructions
Version 3
Valid as of Software V2.0
Adjustable parameters and possible value ranges
Program structure
Error messages
The furnace continuously checks all functions during operation.
If an error is detected, the respective error message is
displayed.
In case of an error, the heater switches off for safety reasons.
Index Category Error
ERR
Conti-
Error Message Text
No.
nuation
possible
1
Entry
T < B
2
Enter a logical value for T
2
Entry
L > T
8
Enter a logical value for long-term cooling L
3
Entry
V2x <= V1x
9
Enter a logical value for the vacuum-on temperature Vx1 or the vacuum-off temperature Vx2
4
Entry
V2x > Tx + 1°C
10
Change either the vacuum values or the holding time T
5
Entry
Incorrect values for V1x, V2x
11
Enter a logical value for V1x, V2x
6
System
Current temperature after
13
Excess temperature! Program aborted, furnace head opens to allow the furnace to cool down.
Start > Tx + 50 °C
*, **
8
Entry
T2 < T1
16
Enter a lower value for T1 or a higher value for T2.
9
System
Power failure > 10 s during
17
A firing program in progress was interrupted for more than 10 s. The program cannot be continued!
a firing program in progress
10
Entry
T1 > V12
18
Enter a lower value for T1 or a higher value for V12
11
Entry
vV set, but V2 is
19
Pre-vacuum activated! V2 must be higher than B.
missing or invalid
12
System
Error in the heating system
20
no
Check the heater fuse. If the fuse is O.K., contact your local Ivoclar Vivadent Service Center.
**
13
System
Heating muffle very old
23
The heating muffle is very old. It is recommended to replace it. After the error message has been
acknowledged, a firing program may still be started.
14
System
Heating muffle defective
24
The condition of the muffle is so poor that is has to be replaced immediately.
16
Entry
T is > B + 200 °C at the
26
Firing chamber too hot to start a firing program.
start of a firing program
17
System
Furnace head cannot
27
The furnace head cannot be moved to the final position. It might be blocked by an external mechanical
be initialized
**, ***
source! If this is not the case, please contact your local Ivoclar Vivadent Service Center!
18
System
The furnace head does not
28
The furnace head does not open/close correctly. The furnace head was manually moved or is obstructed.
reach the target position
**
The furnace head must only be moved using the keys intended for this purpose!
21
System
Necessary vacuum (xxxmbar) 33
The vacuum cannot be established. Check the seal of the firing chamber, vacuum hose, vacuum pump,
is not reached within 1 min.
pump fuse.
33
Entry
HV > H (H2)
110
Enter a lower value for HV or a higher value for H (H2)
35
Entry
"Share of the holding time 120
Activate the vacuum during the holding time Tx or deactivate HV.
with vacuum" is activated,
but Vx2 does not
correspond to Tx or Tx+1
38
System
Brief power failure during a 702
A firing program in progress was interrupted by a brief power failure. The program is continued!
firing program in progress
45
System
Vacuum drop
801
An unacceptable vacuum drop has occurred.
46
System
The vacuum does not
802
No vacuum increase could be measured. Check the following points: Is the firing chamber tight (no
increase (self-test)
contamination on the sealing surfaces)? Is the vacuum hose connected? Is the vacuum pump connected?
Is the fuse F1 o.k.?
75
System
ATK2 calibration:
1302
Error during calibration. Sample may not be correctly inserted. Try again with a new sample and make sure the
Pre-heating to 963 °C
**
sample makes ampl contact.
79
Note
Calibration reminder
1310
Some time has passed since the last calibration procedure. Calibrate the furnace soon.
80
Note
Dehumidification reminder
1312
Some time has passed since the last dehumidification procedure. Dehumidificat the furnace soon.
The following error messages my be displayed:
Please contact the Ivoclar Vivadent After Sales Service, if other error messages being displayed.
Symbol
Parameter
Value range
Value range
P
Program number P
1–20
B
Stand-by temperature
100–700 °C
212–1292 °F
S
Closing time (min : sec)
00:18–30:00
t➚ (*)
Temperature increase rate
30–140 °C/min
54–252 °F/min
T
Holding temperature
100–1200 °C
212–2192 °F
H
Holding time (min : sec)
00.01–60:00
V1
Vacuum on
0 or 1–1200 °C
0 or 34–2192 °F
V2
Vacuum off
0 or 1–1200 °C
0 or 34–2192 °F
t➚ (*)
Temperature increase rate
t2➚
Second stage
30–140 °C/min
54–252 °F/min
T
Holding temperature
Second stage
100–1200 °C
212–2192 °F
H
Holding time
Second stage (min : sec)
00.01–60:00
V1 (V1 2)
Vacuum on
Second stage
0 or 1–1200 °C
0 or 34–2192 °F
V2 (V2 2)
Vacuum off
Second stage
0 or 1–1200 °C
0 or 34–2192 °F
L
Long-term cooling
0 or 50–1200 °C
0 or 122–2192 °F
tL
Cooling temperature rate
0 or 1–50
0 or 32–90
(*) 100 V Version: 140°C/min (252°F/min)
Please observe the notes in the Instructions for Use of the corresponding material.
Program
Description
1
Crystallization/glaze program for IPS e.max CAD
2
Corrective firing for IPS e.max CAD
3
Speed Crystallization/glaze program for IPS e.max CAD
4
Stain and glaze program for IPS Empress CAD
5
Fusion/Crystallization program for the IPS e.max CAD-on technique
6
Characterization/glaze program for the IPS e.max CAD-on technique
7
Crystallization program for IPS e.max CAD Impulse
8 to 20
Individual Programs