beautypg.com

Troubleshooting, 1 low or no reading, Troubleshooting 5.1 low or no reading – Hoefer DQ300 User Manual

Page 26

background image

p18

5. Troubleshooting

Following are troubleshooting tips for problems
the user may encounter.

5.1 Low or No Reading

1. Are all the reagents and standards made to

specification? Are any of the reagents expired?
Is the dilution done correctly?

2. Has the sample been exposed to light for an

extended period?

Some fluorescent dyes are susceptible to photodeg-

radation. It is recommended that the dye solution
be protected from the light.

3. Is the correct optical channel selected for the

application? Has the instrument been calibrated
prior to reading?

4. Is the sample in a detectable, linear range for the

instrument?

Sample concentration might be too low or too high

to be detected (see 4.2 on page 16).

5. Is the right type of cuvette being used?
Methacrylate cuvettes should be used for UV

applications.

6. Does the sample volume meet the minimum

requirement?

The minimum volume is 2 mL for a 10 × 10 mm

cuvette, or 50 µL for a Minicell.

7. Is the a Minicell adaptor inserted in the correct

orientation?

The Minicell adaptor needs to be placed in the

correct position.

If the BLUE channel is selected, insert the Minicell

adaptor in the sample compartment with the tab
away from you and the “BLUE” label facing you.

If the UV channel is selected, insert the Minicell

adaptor in the sample compartment with the tab
nearest to you and the “UV” label facing you.