3.5 Calibration

1

Press the button.

2

Press to start the calibration.

3

Insert your blank and press . The DQ300
will average the fluorescence of the blank for
5 seconds.

4

Insert the calibration standard and press .
The DQ300 will average the fluorescence of the
standard for 5 seconds.

5

Press when the calibration is complete to
accept the calibration. If is not pressed
within 10 seconds, you will be asked if you want to
abort the calibration. Press the or
button to abort or accept the calibration, respectively.


If at any time during steps 1–4 you want to stop
the calibration, press . The display will
return to the Home Screen and default back to
the previous calibration.

Note: If using the Minicell
adaptor, be sure to insert it in
the correct position (see 1.3.5
on page 4).