Guralp Systems CMG-40T User Manual

Page 10

background image

CMG-40T

a DM24, you can do this by sending a

CENTRE command. If you are

using a Handheld Control Unit, you should select 1 SEC VEL from the
Velocity Select switch.

2. Measure the vertical mass position output with a 10 V voltmeter (see

Appendix A, “Connector pinouts”

) or by selecting MASS POS, V from a

Handheld Control Unit's Display Select knob. If using a HCU, also
check that the Centring Select knob is set to OFF.

3. If the vertical component needs adjusting, remove the cap on the lid

which protects the Verticalpot with a flat-bladed screwdriver (provided).

4. Insert the screwdriver through the opening, and engage the pot. An LED

flashlight may be useful for locating the head.

5. Turn the pot either way until the offset readout is as close to 0 V as

possible.

10

Issue A