2 specifications – Fluke Biomedical 190I User Manual

Page 6

background image

Victoreen 1901
Operators Manual

1-2

The four (4) batteries are housed in a separate compartment accessed from the instrument back panel. All four
9-V alkaline batteries are required to operate the unit for approximately 200 hours. Visual and audible low
battery indications are provided whenever battery power is low (four-minute intervals for the audible and
continuous for the visual after a battery test).


To avoid possible damage to the instrument, all
four batteries must be in place.


Communication with the 190I is accomplished through an optional infrared two-way adapter, Victoreen
190I-1A, which mounts at the top of the instrument. The adapter provides positive keying with the
instrument for stable connection and communication.

1.2 Specifications

Radiation Detected

Alpha, Beta, Gamma, or X-Ray depending on detector used

Operating Range

Rate Mode:

CPM: 0-100; 0-1000; 0-10,000; 0-100,000; 0-1,000,000, and 0-

10,000,000

CPS: 0-1, 0-10, 0-100, 0-1000, 0-10,000, 0-100,000, and 0-166,666

μR/hr, mR/hr, R/hr, μSv/hr, mSv/hr, Gy/hr, & Bq/cm

2

: depends on

conversion factor of the probe used.

Integrate Mode:

Counts: 0-100,000,000; displayed as 4 digits up to 9999 counts, 3 digits
plus a K (for thousand) or an M (for Million) at higher counts.

Seconds: 0-65535; displayed as 4 digits up to 9999 seconds, 3 digits in
minutes and tenths of minutes above 9999 seconds.

μR, mR, R, cts, μSV, & mSv: depends on conversion factor of the probe

used.

Accuracy

Within 10% of reading between 10% and 100% of full-scale indication on
any range, exclusive of energy response.

Ranges

Six (6) decades are available for calibration; the number of usable
decades is detector dependent.

Deadtime Correction

Provided above 1000 CPM

Jam Detection

Adjustable level for determining jam condition; provided in probe adapter
modules.

Alarm

Can be set on any decade at 10% to 100% (in increments of 10%) of full

scale.

Detector

Internal Geiger-Mueller Tube

Warm-up Time

15 seconds (diagnostic testing)

Response Time (Bargraph) Selectable; 24, 12, 6, and 3 seconds.

24-second response time selected:

Scale Range

(CPM) Response

(seconds)

1

0-100

24.0

2

0-1000

12.0

3

0-10K

3.0

CAUTION