Page 6

background image

943-27 Beta Scintillation Detector
Operator Manual

1-2

1.3 Specifications

General specifications for the Model 943-27 Detector are listed below. The 943-27 series products are
designed for nuclear applications, and any repairs to it by personnel not qualified to ANSI 45.2.6 1978,
Skill Level II may void the nuclear rating.

If a problem develops, the preamplifier may either be returned to the factory for service, or repaired by a
qualified technician.

Table 1-2. General Specifications for the 943-27 Detector

Feature: Specification:

Radiation Detected

Beta rays

End Window

Titanium, 1.125 in dia. x 0.0002 in thick, designed to maximize beta
transmission and minimize noble gas adsorption

Scintillation Crystal

Calcium Fluoride, CaF

2

(Eu)

1.125 in dia. X 0.005 in thick interfaced to 1.125 in dia. x 0.125 in thick
quartz light pipe

Detector Housing

Stainless Steel

Detector Constant

1.0 x 10

-3

uCi/cc/picoampere, typical, based on calibration method and

sampling geometry

Detector Range

1.0 x 10

-12

to 1.0 x 10-4 amperes (approximately 1.0 x 10

-3

to 1.0 x 10

5

uCi/cc, based on calibration method and sampling geometry

Energy Dependence
and Efficiency

Dependent on sampling geometry; Refer to Appendix C for applicable
isotopic calibration reports

Power Requirements

(-) 500 VDC @ 400 uAmperes

Operating Temperature

0° F to 122 °F (-16 °C to 50 °C)

Relative Humidity

0 – 95 %, non-condensing

Operating Pressure

12 psig max

Dimensions

1.5 in dia. x 9.0 in long (3.8 cm x 22.9 cm)

Weight

3 lbs (1.4 kg)

Preamplifier Interface

High Voltage: Six (6) foot coaxial cable with pre-terminated SHV connector
and “Seal Grip”;
Signal: Six (6) foot coaxial cable with pre-terminated BNC connector and
“Seal Grip”;

Modification
Information

None