3 troubleshooting – Fluke Biomedical 943-27 Detector User Manual
Page 23

Maintenance, Calibration and Troubleshooting
Troubleshooting
4
4-9
4.3 Troubleshooting
WARNING
Extreme care must be used when troubleshooting a 
system that has power applied. All standard 
troubleshooting precautions apply. 
WARNING
Once a problem has been located, remove all power 
before continuing with the repair. 
CAUTION
Personnel performing the troubleshooting/repair 
must be qualified to ANSI 45.2.6, 1978, Skill Level II. 
Personnel performing the following procedure must 
be familiar with the operation of the monitoring 
system and the location of each piece of equipment 
used in the system. 
 
If a problem develops, verify that the voltages at connection point inputs and outputs are present and that 
all wiring is secure. Refer to Appendix B and C for drawings and parts lists. If a PROM requires 
replacement, specify the board revision level when ordering the part. 
 
 
 
Troubleshooting / Adjustments
 
Troubleshooting includes a verification of detector response by actuating the radioactive check source 
and verification of the high voltage value. The check source is actuated from the associated 
readout/controller. The high voltage may be measured at the connection point in the digital preamplifier 
enclosure. 
 
Specialized test equipment and firmware are required to adjust the auto-zero and PGA subsystems in the 
electrometer. If the test equipment and firmware is not available, the unit should be returned to Fluke 
Biomedical RMS for re-alignment. 
 
 
 
