Fluke Biomedical 943-27 Detector User Manual
Page 20

943-27 Beta Scintillation Detector
Operator Manual
4-6
Source Holder Data Sheet:
The following summarizes the data provided with the source holders:
CURRENT MODE BETA SOURCE SET
P/N 844-130-1-12
90
Sr
P/N 844-130-1-13
137
Cs
P/N 844-130-1-14
36
Cl
USE PRECAUTIONS FOR HANDLING SOURCES
RADIOACTIVE MATERIAL
WARNING: HANDLE WITH CAUTION
HIGH BETA RADIATION FIELDS EXIST WHEN END CAP IS REMOVED
DIRECT THE OPEN SOURCE AWAY FROM PEOPLE
WEAR SAFETY GLASSES WHEN USING THE SOURCES
DO NOT TOUCH THE SOURCE WINDOW – THE THIN WINDOW WILL
PUNCTURE EASILY
U. S. NUCLEAR REGULATORY COMMISSION REQUIRES SIX MONTH LEAK TESTING
NOTICE: THE LARGE END CAP CONTAINS THE RADIOACTIVE SOURCE
CAPSULE
DO NOT REMOVE
PART
NUMBER
SERIAL
NUMBER,
FLUKE
SERIAL
NUMBER,
MANUFACTURER
RADIOACTIVE
MATERIAL
ACTIVITY,
microcuries
DATE OF
MEASUREMENT
844-130-1-12 XXXX
XXXX
90
Sr
200 XX/XX/XX
844-130-1-13 XXXX
XXXX
137
Cs
200 XX/XX/XX
844-130-1-14 XXXX
XXXX
36
Cl
200 XX/XX/XX
Detector Calibration Procedure
The following procedure is recommended to assure an accurate calibration of the Model 943-27 Current
Mode Beta scintillation detector.
Equipment Required:
1. Detector to be calibrated
2. Electronics to be used with the detector
3. Calibration sources P/N 844-130-1-12, 844-1-130-1-13 and 844-130-1-14.
4. Data Sheet: The data sheets are used to delineate the minimum recording requirements. The
baseline calibration data is supplied at the time of installation. Record all identification
information on the data sheet prior to doing any measurements. The expected calibration
values must be corrected for decay prior to doing the calibration. Refer to the calibration
procedures in Appendix B for sample data sheets.