943-27 Beta Scintillation Detector
Operator Manual

4-6


Source Holder Data Sheet:
The following summarizes the data provided with the source holders:

CURRENT MODE BETA SOURCE SET

P/N 844-130-1-12

90

Sr

P/N 844-130-1-13

137

Cs

P/N 844-130-1-14

36

Cl

USE PRECAUTIONS FOR HANDLING SOURCES

RADIOACTIVE MATERIAL

WARNING: HANDLE WITH CAUTION

HIGH BETA RADIATION FIELDS EXIST WHEN END CAP IS REMOVED

DIRECT THE OPEN SOURCE AWAY FROM PEOPLE

WEAR SAFETY GLASSES WHEN USING THE SOURCES

DO NOT TOUCH THE SOURCE WINDOW – THE THIN WINDOW WILL

PUNCTURE EASILY

U. S. NUCLEAR REGULATORY COMMISSION REQUIRES SIX MONTH LEAK TESTING

NOTICE: THE LARGE END CAP CONTAINS THE RADIOACTIVE SOURCE

CAPSULE

DO NOT REMOVE

PART

NUMBER

SERIAL

NUMBER,

FLUKE

SERIAL

NUMBER,

MANUFACTURER

RADIOACTIVE

MATERIAL

ACTIVITY,

microcuries

DATE OF

MEASUREMENT

844-130-1-12 XXXX

XXXX

90

Sr

200 XX/XX/XX

844-130-1-13 XXXX

XXXX

137

Cs

200 XX/XX/XX

844-130-1-14 XXXX

XXXX

36

Cl

200 XX/XX/XX

Detector Calibration Procedure

The following procedure is recommended to assure an accurate calibration of the Model 943-27 Current
Mode Beta scintillation detector.

Equipment Required:

1. Detector to be calibrated
2. Electronics to be used with the detector
3. Calibration sources P/N 844-130-1-12, 844-1-130-1-13 and 844-130-1-14.
4. Data Sheet: The data sheets are used to delineate the minimum recording requirements. The

baseline calibration data is supplied at the time of installation. Record all identification
information on the data sheet prior to doing any measurements. The expected calibration
values must be corrected for decay prior to doing the calibration. Refer to the calibration
procedures in Appendix B for sample data sheets.