Maintenance, calibration and troubleshooting, 1 maintenance, 2 calibration – Fluke Biomedical 943-27 Detector User Manual
Page 15

Maintenance, Calibration and Troubleshooting
Maintenance
4
4-1
Section 4
Maintenance, Calibration and Troubleshooting
4.1 Maintenance
The 943-27-15 Detector is designed to operate for extended periods of time with no scheduled 
maintenance required. Operation may be verified by periodically actuating the check source (provided 
with the detector and sampling assembly) and observing the response of the unit. If the response varies 
by more than 50% of the normal value, further troubleshooting may be required. If a problem develops, 
troubleshoot the unit per Section 4.3 and the drawings in Appendix B. 
 
Periodically, approximately every 18 months, it is suggested that the detector be re-calibrated. 
 
 
4.2 Calibration
Electronic Adjustments
There are no user adjustments available on the 943-27 Detector. The detector high voltage is determined 
during calibration and is provided by the associated readout/controller. 
 
Point source Calibration
WARNING
The calibration procedures require the use of high activity 
beta emitting button sources. These sources are intended 
solely to provide a beta radiation field. They should be 
used only by persons who have been trained in the 
appropriate safety procedures to be followed in the 
presence of beta radiation and the proper interpretation of 
the data. Notify the plant Health Physicist prior to the use 
of these sources. 
 
 
SPECIAL NOTE
Due to the activity of these sources, possession of these 
sources requires a USNRC Byproduct Material License or 
Agreement State License, as defined in Title 10 of the 
Code of Federal Regulations. 
