Fluke Biomedical 875 Victoreen User Manual
Page 7

Victoreen 875 High Range Containment Monitor
Operator Manual
1-2
Hermetic Integrity
1E10-5
cm
3
of helium per second @ one standard
Chamber Fill Gas
(2%) at tmospheric pressure
ensions (W x H x D) with
acket
t
Approximately 18 lb (8.17 kg)
emperature, Operating
1 °C)
375 °F (180
ccident, 3 hours
elative Humidity
100 % (wate
otal Integrated Dose
2E8 Rads + 1
60
Co
, 24 hrs
28 days
e description of Qualification Test Results)
A
+36 % of input radiation
ut radiation
a) AC Voltage: 108 to 132 VAC, RMS @ 60
± 3 Hz
21 in
(214.88 x 133.35 x 386.33 mm)
°C to 60 °C)
F (4 °C to 49 °C)
Relative Humidity (Storage): 0 % to 95 % (non-
condensation)
atmosphere
Nitrogen/helium
a
Dim
9.25 x 12.5 x 10.06 in
Mounting Br
(234.95 x 317.5 x 255.52 mm)
Weigh
Temperature, Storage
40 °F to 160 °F (4 °C to 71 °C)
T
40 °F to 160 °F (4 °C to 7
°C) peak, a
R
rproof)
T
0 % margin,
Chemical Spray
0.45 Gallons/minute/ft
2
(0.28 molar H
3
BO
3
, 3000 ppm
Boron, and NaOH; ph of 11
LOCA Test Duration
(Consult Fluke Biomedical for a complet
Readout Module 876A-1
Range
1 to 10
7
R/h
Input Current
Minimum: 6.5 to 7.5 x 10
-11
A
Maximum: 6.5 to 7.5 x 10
-4
Recorder Output
0 to +1 VDC, logarithmic (0 - 10 mV and 0 – 5 V
optional)
Computer Output
0 to +5 VDC, logarithmic (0
-
100 mV and 0
-
50 mV
optional)
(Other Outputs Available Upon Request)
System Accuracy (during "all" conditions)
Accumulative @ Meter
Analog Outputs +28 % of inp
Power Requirements:
b) Battery Power: 22 to 32 VDC @ 600 mA DC
maximum
Temperature Coefficient
0.40 %/°C, + 0.25 R/h/°C
Dimensions (W x H x D)
8.46 x 5.25 x 15.
Weight
20 lbs (9.07 kg)
Environmental Parameters
Temperature (Storage): 40 °F to 140 °F (4
Temperature (Operating): 40 °F to 120 °