Fluke Biomedical 875 Victoreen User Manual
Page 33

Victoreen 875 High Range Containment Monitor
Operator Manual
5-2
NOTE
Detector 877-1 from Readout Module
t Equipment
Disconnect
876A-1 by removing cables from J1 and J2 before
starting electronic calibration.
Table 5-1. Recommended Tes
Current Generator - Test Electrometer
Range
1 x 10
-12
Amps to 1 x 10
-3
Amps
Accuracy, Current Generator
10
-12
Amps
± 10 %
10
-11
Amps
± 3 %
-10
-7
10
to 10 Amps
± 2 %
10
-7
to 10
-3
Amps
± 0.2 %
Accuracy Test Electrometer
± 0.2 % ± 1 digit with integration time long enough to accumulate 3
significant digits of volts indicated on display V/R
± 0.2 % ± 1 digit
Adjustable AC Voltage Source
Maximum Voltage
Low range 140 VAC, high-range 280 VAC
Frequency 60
Hz
4 ½ Digit Digital Multimeter
Fluke 8600A or equivalent
Ranges
+ 0 to 100 mV up to + 0 to 100 V (20 % over-range)
Accuracy
100 mV range:
± (0.05 % of input ± 0.02 % of range)
all other ranges:
± (0.02 % of input ± 0.01 % of range)
In
put Impedance
> 10 megohms
D
C Voltage
± 0.03 %
AC Voltage
± 1 %
DC Current
± 0.3 %
Trim Pot Adjustment Tool
Carbon Composition Resistor
Resistance
10 kilohms, ½ W, 5 %
Lead Connectors
Small alligator clips
DC Power Supply
Output Voltage
0 to + 25 VDC
Output Current
0 to 600 mA
L
0.01 %
± 4 mV
oad Regulation
A
it
Down to 10 mA
djustable Current Lim
M
nter
ulti-function Cou
F
luke 1900A or equivalent
Frequency
5 Hz to 80 MHz (0.1 to 100 Hz Resolution)
Period
5 Hz to 1 MHz, single and multiple period averages
(1 to 100 nsec resolution)
Total Counts
1 to 999,999 counts