Fluke Biomedical 875 Victoreen User Manual
Page 43

Victoreen 875 High Range Containment Monitor
Operator Manual
5-12
able 5-1. Truth Table for Pass and Fail Conditions in The Latch Circuit
(Circuit assumed initially in Pass Condition)
T
Fail Condition
Pass Condition
Pin
Voltage
State
Voltage Comments
Pin
State
Comments
4
L (low)
Since 6 is H (high)
4
H
Because 6 is L (also 5)
3
H (high)
Since 1 & 2 are Low
3
L
Because 2 is H
13
H
Connected to 3
8
H
Connected to 4
11
L
Since 13 is H
13
L
Connected to 3
9
L
Connected to 11
10
L
Because 8 is H
8
L
12
L
Connected to 4
10 H
9, 10) is
11
H
Bec
Since U202, (8,
NOR
ause 12 and 13 are L
12 H
from Lo
Result: Output switches
w to High
This is the original condition. No
change takes place.
Notes:
ition assumes 12 and 1
The truth table shows that no change in the output can occur without the presence of the enable pulse,
the monitoring signal indicat
the ramp voltage
able 5-2. Signal In Fail State (High Signal Input)
Pin 1
Voltage State
Comments
1) Enable pulse assumed present, so 1 and 5 are L.
2) Initial cond
0 L.
even though
es a fail condition. This is necessary because
starts from zero during the ECS Test.
T
10, U208 and 6, U202
H
Fail state
9, U208 and 2, U202
L
After inversion
1,5, U202
H
No enable input
4, 8, U202
L
Because of NOR action
3, 13, U202
L
Because of NOR action
10, 12, U202
L
Because of NOR action
Thus the circuit does not react to a fail signal input unless pins 1 and 5 are low.
If a problem cannot be resolved by applying the
troubleshooting and maintenance procedures
provided here, please contact Fluke Biomedical at
440.248.9300 for assistance.