Fluke Biomedical ESA620 User Manual
Page 75

Electrical Safety Analyzer
Detailed Specifications
61
Insulation resistance
Ranges ................................................................ 0.5 to 20 M
Ω
20 to 100 M
Ω
Accuracy
20 M
Ω
Range ..................................................
±
(2 % of reading + 2 counts)
100 M
Ω
Range ................................................
±
(7.5 % of reading + 2 counts)
Source test voltage ............................................. 500 V dc (+20 %, -0 %) 1.5 mA short-circuit current or 250 V dc selectable
Maximum load capacitance ................................. 1
μ
F
ECG Performance Waveforms
Accuracy..............................................................
±
2 %
±
5 % for amplitude of 2 Hz square wave only, fixed @ 1 mV Lead II configuration
Waveforms
ECG Complex ................................................. 30, 60, 120, 180, and 240 BPM
Ventricular Fibrillation
Square wave (50 % duty cycle) ....................... 0.125 and 2
Sine wave ........................................................ 10, 40, 50, 60, and 100 Hz
Triangle wave .................................................. 2 Hz
Pulse (63 ms pulse width ................................ 30 and 60