Electrical Safety Analyzer
Detailed Specifications
59
Additional error caused by series inductance
Resistance
Series Inductance
0 µH
100 µH
200 µH
400 µH
0.000
Ω
0.000
Ω
0.030
Ω
0.040
Ω
0.050
Ω
0.020
Ω
0.000
Ω
0.025
Ω
0.030
Ω
0.040
Ω
0.040
Ω
0.000
Ω
0.020
Ω
0.025
Ω
0.030
Ω
0.060
Ω
0.000
Ω
0.015
Ω
0.020
Ω
0.025
Ω
0.080
Ω
0.000
Ω
0.010
Ω
0.015
Ω
0.020
Ω
0.100
Ω
0.000
Ω
0.010
Ω
0.010
Ω
0.015
Ω
>0.100
Ω
0.000
Ω
0.010
Ω
0.010
Ω
0.010
Ω
Equipment Current
Range .................................................................. 0 – 20 A ac rms
Accuracy.............................................................. 5 % of reading
±
(2 counts or 0.2A, whichever is greater)
Duty cycle ............................................................ 15 A to 20 A, 5 min. on/5 min. off
10 A to 15 A, 7 min. on/3 min. off
0 A to 10 A continuous
Leakage Current
Modes* ................................................................ AC+DC (True-rms)
AC only
DC only
* Modes: AC+DC, AC only, and DC only available for all leakages with exception of MAP
that are available in True RMS (shown as AC+DC)
Patient Load Selection ........................................ AAMI ES1-1993 Fig. 1
IEC 60601: Fig 15
IEC 61010: Fig A-1