KV Correction for Various Target/Filter Combinations
D
D-1
Appendix D
kV Correction for Various Target/Filter Combinations
D-1 kV Correction for Various Target/Filter Combinations
The following tables and graphs may be used to correct 4000M+ kVp readings taken with the following
Mammographic x-ray tube target/filter combinations:
Molybdenum target 30µ Molybdenum filter
Molybdenum target 25µ Rhodium filter
Molybdenum target 1 mm Aluminum filter
Rhodium target
25µ Rhodium filter
Rhodium target
1 mm Aluminum filter
Two forms of correction have been calculated, one is a multiplier and the other is a kV offset that may be
added to the indicated kV. These tables’ ad graphs are only valid when a 4000M+ is used in the
Mammographic mode (Mo/Mo selected) with the standard Mo/Mo calibration. The use of these factors
will correct the indicated kV to 1 kV over the Mammographic kV range.
- 7000DP Impulse Supplement (12 pages)
- 7000DP Impulse Getting Started (36 pages)
- 7000DP Impulse (92 pages)
- 06-526-2200 (26 pages)
- 07-417 (16 pages)
- 07-443 (14 pages)
- 07-451 (8 pages)
- 07-487 (12 pages)
- 07-453 (24 pages)
- 07-555 (8 pages)
- 07-553 (10 pages)
- 07-605-7777 (8 pages)
- 07-444 (46 pages)
- 07-618 (10 pages)
- 07-600 (6 pages)
- 07-591 (8 pages)
- SigmaPace 1000 (154 pages)
- 07-653 (8 pages)
- 07-633 (20 pages)
- 07-649 (14 pages)
- 07-661-7662 (12 pages)
- 07-645 (14 pages)
- 10100AT (80 pages)
- 07-644 (10 pages)
- 18-203 (4 pages)
- 07-621 (12 pages)
- 07-647 (12 pages)
- 18-207 (8 pages)
- 18-216-1000 (10 pages)
- 18-220 (10 pages)
- 18-228 (8 pages)
- 18-223 (18 pages)
- 18-229-1313 (6 pages)
- 18-250 (8 pages)
- 18-252 (6 pages)
- 18-303 (6 pages)
- 35035 (19 pages)
- 6000-528 (18 pages)
- 35080B (70 pages)
- 6000-529 (16 pages)
- 35080M (64 pages)
- 6000-530B (16 pages)
- 57-436 (12 pages)
- 57-440 (12 pages)
- 57-402 (32 pages)