beautypg.com

Floscan instrument company, inc – Floscan 5A20-20A-1 AccuTroll & CruiseMaster User Manual

Page 21

background image



08/20/2004

4001-102-00I

FloScan Instrument Company, Inc.

Tel: (206) 524-6625

Fax: (206) 523-4961

3016 NE Blakeley Street, Seattle, WA 98105

Email:

[email protected]

Http://www.floscan.com

4. Move the voltmeters’ negative lead to the BROWN (Signal) wire. With engine idling, measure and record

the voltage reading. Readings between 4 and 10 VDC, usually indicate a good sensor.

___________VDC

5. Stop the engine and observe the voltmeter. Voltage readings should fluctuate between a high of 9 to 12

VDC, and a low of 0 to 4 VDC as the Sensors’ turbine slows to a stop.

NOTE: This may not be seen with a digital voltmeter.

6. The Sensor may be defective if voltage readings in steps 3 & 5 remain constant while the engine is

stopping.

7. On twin engine systems, try swapping the Brown & White signal wires. Observe the Instrument and see if the

problem shifts sides, or remains on the same side.

With this guide in hand, contact FloScan Technical Support.

III. BACKLIGHTING.


1. Referring to the wiring diagram, check back lighting bulb and Instrument illumination circuit.


IV. CHECK + 12 VDC and GROUND

1. Turn Instrument power ON, and go to the Instruments’ backside.

2. Measure between the RED (Power) wire and the BLACK (Instrument ground) wire.

__________VDC

This reading should be approximately 12 VDC, but not lower than 10 VDC.

3. If 12 VDC is not present, measure between the RED power wire and a known good ground in the

instrument panel. If you measure 12 to 14 VDC between the Red wire and ground, you may have

__________VDC

an instrument ground problem.

4. If 12 to 14 VDC is not present in steps 2 or 3, check wiring, switches, fuse, and the 12 VDC power source.


V. CONTINUITY TESTING

1. Continuity testing requires access to the back of the Instrument and an Ohmmeter. It verifies that wires are not

broken, shorted to ground, another wire, or to power. Before starting, secure all power.

2. Disconnect the WHITE and BLACK wire harness conductors from the WHITE and BLACK Port Sensor

wires. Connect a jumper from the WHITE wiring harness conductor to a known good ground. At the

____________

back of the Instrument, connect an Ohmmeter lead to its’ WHITE wire. Connect the other Ohmmeter lead

(Ohm Reading)

to a known, good ground. The Ohmmeter should read approximately 0.1 to 3

Ω.

3. Disconnect the BROWN and BLACK wire harness conductors from the WHITE and BLACK Starboard Sensor

____________

wires. Connect a jumper from the BROWN wiring harness conductor to a known good ground. At the back of the

(Ohm Reading)

Instrument, connect an Ohmmeter lead to its’ BROWN wire. Connect the other Ohmmeter lead to a known good
ground. The Ohmmeter should read approximately 0.1 to 3

Ω.


VI. PULSATION PROBLEMS

A stiff anti-siphon valve, sticking carburetor float or miss-plumbed Pulsation Damper(s) can cause fluctuating GPH readings. These
are most pronounced at low Rpm’s and tend to disappear around mid throttle. They are annoying, but won’t affect totalizer accuracy.
To eliminate, repair or replace any defective parts, or re-plumb dampers according to the Fuel Flow Schematic Instructions.

This manual is related to the following products: