SF100 Filtration Unit

2.5 Liquid Connections

The SF100 Self Cleaning Filter Unit is designed for vertical mounting with

sample flow moving from the bottom to the top of the assembly. This

allows the air bubbles generated in the cleaning cycle to easily purge

from the system.

The 1” FNPT process connections are pipe union mounted to the Flow

Assembly. Allow 8-10” around the fitting for easy access during

maintanence. A sample shut off valve is recommended below the flow

assembly to facilitate maintanence. The filter assembly should be

mounted at a similar height to the CA-6 Analyzer Fast Loop Reservoir

sample port.

The optional SF100 Mounting Kit (PN 1000006) is available containing

Schedule 80 PVC pre-cut piping, fittings, ball valve and the mounting

brackets as shown in the drawing. The overall assembly is 7.5’ high

with 1” FNPT input and 1 ½” FNPT outfall.

Sample pressure should be no higher than 35 psi. The sample

pressure will dictate the flow rate though the filter. Care should be

taken since a high sample line pressure may lead to a high flow rate

by the Filter Assembly generating a venturi effect thereby negating

the line pressure.

The sample out fitting is a ¼” tube fitting. Flexible or hard tubing can

be used. The outlet should be run to a Fast Loop Reservoir which

provides an atmospheric sample point for the CA-6 Analyzer.