SF100 Filtration Unit

1.0 INTRODUCTION

1.1 System Description

The SF100 Self Cleaning Filter Unit is designed to protect the

CA-6 Colorimetric Analyzer from fouling due to particles in the

sample stream. Air pressure is used to clean the filter element.

Timers periodically actuate a solenoid that sends pressurized air

back through the filter element, blowing any material off of the

filter and back into the sample line. Both the period and the

duration of the Blow Back Cycle are user selectable.

The Filter Unit should be plumbed into an area of high velocity

to maximize the time between cleaning cycles. Higher velocities

seem to minimize sedimentation on and fouling of the filter

element.

1.2 Specifications

Filter Mesh:

100 micron Standard (50-1000 micron upon request)

Materials of Construction:

Body-Polypropylene

Filter- INOS AISI 316 Stainless Steel

Solenoid Valve- INOS AISI 316 Stainless Steel and BUNA

Temperature Range:

35°-131°F (2°- 55°C)

Sample Line Pressure:

Minimum 5 psig and Maximum 35 psig (0.3-2.5 bar)

Sample Flow Rate:

Minimum 25 gallons/hour (0.1m

3

/hour)

Air Pressure:

At least 20% above sample line pressure, 45 psig max (3 bar)

Filtered Sample Flow Rate:

0.1 to 2.0 liters/minute, depending on line pressure

Flow Cell Connection:

1” FNPT/Pipe Union Connection

Filter Connection (Air):

6 mm Tube

Filter Connection (Sample Out):⅟₄” Tube

Power Supply:

220 VAC (110 VAC optional)

Consumption 20VA

Protection Rating:

IP65

Cleaning Interval:

1-45 minutes (user selectable)

Blowback Interval:

1-30 seconds (user selectable)

Weight:

2.2 pounds (1kg)