10.0 ANALYZER SHUT DOWN
If the CA-6 Analyzer will be out of service for a period of two weeks or greater, proceed as follows:
1.
Empty all reagent containers.
2.
Rinse and refill all reagent containers with distilled water.
3.
Prime all of the pumps with DI water. (login with 1
st
level password, press DISPLAY,
MANUAL STEP, ADD REAG 1,2,3,4 for 40 seconds each)
4.
Disconnect the sample feed line and fill the fast-loop reservoir (if present) with distilled
water.
5.
With the sample inlet tubing attached to a container of distilled water, run the analyzer
for at least 2 cycles.
6.
Empty the water from all lines.
7.
Put the analyzer in stand-by condition.
8.
Turn OFF the power to the analyzer and disconnect the plug from the wall socket.