3 accessories and spare parts – Electro-Chemical Devices (ECD) CA6 Hardness Analyzer User Manual
Page 72
9. Position the tubing and fittings deep into the pump
head body.
10. Using Manual Mode, activate the peristaltic pump for 1
second and insert one roller at a time in the free
position (where the roller is not in contact with the
tubing). Repeat for the 3 rollers.
11. Replace the Plexiglas pump cover.
Attach the cover
with the 3 screws TSPEI M3x8 using an Allen Key.
12. Using Manual mode activate the peristaltic pump for
180 seconds (3 minutes) to settle the tubing into
position.
13. Reconnect the outlet tubing to the peristaltic pump fitting.
14. Reconnect the reagent intake tubing to the peristaltic pump fitting.
15. Using Manual mode activate the peristaltic pump for 45 seconds to prime the pump.
16. Repeat the above steps for each additional pump for which it is necessary.
Maintenance frequency for tubing replacement is 70 working hours. To calculate the maintenance
frequency for a pump, simply determine the number of analysis cycles per day and the reagent's dose
time for each analysis cycle.
Example: 4 analysis cycles per hour (i.e. 96 analysis cycles per day), Dose time 1 5 seconds for each
analysis cycle. Working hours per day = (96 cycles/day)(15 sec) / 3600 sec/hour = 0.4 hours / day
Maintenance frequency = 70 hours / 0.4 hours / day = 175 days
9.3 Accessories and Spare Parts
Model CA6 Analyzer: Spare Parts
Part#
Description
2000005
*Recommended Spare Part
Yearly Maintenance Kit, Norprene size #16 tubing (4 each), Pinch Valve Tubing (6 each),
Colorimetric Cell O-Rings (2 each) Tubing Replacement Tool (Sample Pump)
2010036-1 Hardness Analyzer, 2 Reagent Kit, 2 x 500 ml bottles
2010037-1 Hardness Reagent #1 , 500 ml bottle
2010038-1 Hardness Reagent #2 , 500 ml bottle
2010039-1 Calibration Solution, Hardness, 100 ppm, 2 liter bottle
9380012
Mixing Pump - PP EPDM
9380013
Mixing Pump - PP Teflon
9370044
Pinch Valve
9370045
3-Way Valve
3000004
Support for 3-Way Valve
8700001
Heater Element (Thermostatic Block)
3000001
Quartz Reaction Cell 16 mm *Recommended Spare Part
61