8.6 PROGRAM MENU
Menu Functions:
Analysis Cycle
Allows the user to program the 30 steps available in the
Analysis Cycle. This function is protected with the 2
nd
level password.
Extra Cycle
Allows the user to program the 30 steps available in the
Extra Cycle. This function is protected with the 2
nd
level
password.
Settings
Set the ratio of the Analysis Cycles to Extra Cycles, the
Alarm Value (low or high), the 4-20 mA Output and to
Enable or Disable the Loss of Sample Alarm.
Calibration Menu
Allows the user to perform Blank and Factor
Calibrations and to enter the value of the Standard
Solution used for calibration. This function is protected
with the 1
st
level password.
8.6.1 Analysis Cycle
Analysis Cycle displays the numerical
step, the time (in seconds) and the
function for each of the 30 available
steps in the Analysis Program. Touching
a function changes it to the next
function in the list; all functions are
available in this circular list.
37