Digi-Star EZ 2000/V User Manual
Page 11

Digi-Star Model EZ2000/EZ2000V
3. Mounting and connection
9
LIGHTNING PROTECTION
Additional protection can be achieved with
the proper installation of grounding rods.
Please call (920)
563-9700 and request
Digi-Star Form F3050.
TECHNICAL MANUAL
Technical Manual available upon request.
Please call (920) 563-9700 and request
Digi-Star Manual for New EZ2.
CALIBRATION
Warning!
This indicator was calibrated at the factory
to weigh accurately with your system.
Additional calibration is not necessary under
normal conditions.
The Short Form Setup & Calibration
procedure allows you to change the
“SETUP” and “CAL” numbers of the
indicator. You may want to perform this
procedure if;
1. The indicator is being connected to
different load cells, or
2. You want to adjust the calibration to
match another scale system.
Before continuing, first write down the
current SETUP and CAL numbers of your EZ
indicator. These numbers are displayed
during the Self Test.
To run the self test:
With the indicator already ON, press the
key to start the Self Test. Press the
key
to "pause" the Self Test while numbers are
displayed. Press
again to "resume".
SETUP # ___________ CAL # ____________
Keep this information for future reference.
!
Do not attempt to calibrate the scale if
the indicator is not reading stable weights.
The calibration procedure will not fix
instability, inconsistencies, or flashing
"RANGE" messages.