Important settings:
For fast and accurate proportional dosing it
is import to create a small overshoot, like
10% as shown on the left. By doing this the
Coriolis instrument has more time to control
the requested dosage.
This is especially important for small
dosages. The graph shows a dosage of 20
gram in 5.5 seconds. Please see the correct
settings on the next page.
An overshoot can be created by increasing
the PID-Kp and PID-Ti value.
8.4
FlowPlot settings
These settings are used to dose a batch of 20 gram.