Beurer BC 19 User Manual
Page 65

65
Table 3
For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacturer’s declaration – electromagnetic immunity
The BC 19 is intended for use in the electromagnetic environment specified below. The customer or the user of the BC 19 should
assure that it is used in such an environment.
Immunity test IEC 60601 test
level
Compliance
level
Electromagnetic environment – guidance
Radiated RF
IEC 61000-4-3
3 V/m 80 MHz
to 2.5 GHz
3 V/m
Portable and mobile RF communications equipment should be used no closer to
any part of the BC 19, including cables, than the recommended separation dis-
tance calculated from the equation applicable to the frequency of the transmitter.
Recommended separation distance:
d = 1.2
80 MHz to 800 MHz
d = 2.3
800 MHz to 2.5 GHz
Where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic
site survey,
a
should be less than the compliance level in each frequency range.
b
Interference may occur in the vicinity of equipment marked with the
following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.