Beurer BC 19 User Manual
Page 64

64
Table 2
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacturer’s declaration – electromagnetic immunity
The BC 19 is intended for use in the electromagnetic environment specified below. The customer or the user of the BC 19 should
assure that it is used in such an environment.
Immunity test
IEC 60601 test level Compliance level Electromagnetic environment – guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or ceramic tile. If
floors are covered with synthetic material, the relative
humidity should be at least 30 %.
Power frequency (50 / 60 Hz)
magnetic field IEC 61000-4-8 3 A/m
3 A/m
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical com-
mercial or hospital environment.
NOTE: U
T
is the a.c. mains voltage prior to application of the test level.
- HK 63 (52 pages)
- HK Comfor (20 pages)
- HK 49 Cosy (52 pages)
- HK 125 XXL (56 pages)
- FWM 40 (44 pages)
- FWM 50 (44 pages)
- HD 90 (52 pages)
- FB 50 (44 pages)
- FB 50 (48 pages)
- СM50 (48 pages)
- MG 18 (36 pages)
- MG 18 (36 pages)
- MG 18 (92 pages)
- MG 150 (36 pages)
- MC 3000 (84 pages)
- MG 158 (44 pages)
- MG 170 (44 pages)
- MC 5000 (108 pages)
- MG 310 (60 pages)
- MG 240 (56 pages)
- MG 310 (60 pages)
- MG 310 (60 pages)
- MG 310 (52 pages)
- MG 310 (52 pages)
- EM 35 (60 pages)
- EM 35 (60 pages)
- EM 41.1 (112 pages)
- EM 38 (74 pages)
- BM 70 (100 pages)
- FW 20 (48 pages)
- HK 48 Cosy (56 pages)
- HK 65 Aroma (28 pages)
- UB 90 Pure (60 pages)
- UB 90 Pure (72 pages)
- HK 72 (60 pages)
- UB 85 (44 pages)
- BC 16 (88 pages)
- BC 20 (68 pages)
- BC 40 (104 pages)
- BC 30 (80 pages)
- BC 31 (92 pages)
- BC 32 (76 pages)
- BC 44 (84 pages)
- BC 44 (44 pages)