8. Cleaning and storing the instrument
• Clean your blood pressure monitor carefully using a
slightly damp cloth only.
• Do not use any detergents or solvents.
• Never hold the instrument under water as otherwise
liquid can penetrate and damage the instrument.
• Never place any heavy objects on the instrument.
9. Technical details
Model no.
BC 19
Measurement
method
Oscillometric, non-invasive blood
pressure measurement on the
wrist
Measurement range
Cuff pressure 0 – 300 mmHg,
systolic 60 – 260 mmHg,
diastolic 40 – 199 mmHg,
Pulse 40 –180 beats/minute
Display accuracy
Systolic ± 3 mmHg,
diastolic ± 3 mmHg,
pulse ± 5 % of the value shown
Measurement inac-
curacy
Max. permissible standard devia-
tion according to clinical testing:
systolic 8 mmHg /diastolic
8 mmHg
Memory
2 x 60 memory spaces
Dimensions
L 85 mm x W 70 mm x H 31 mm
Weight
Approx. 140 g (without batteries)
Cuff size
14.0 to 19.5 mm
Permissible operating
conditions
+10 °C to +40 °C, ≤ 85 % relative
air humidity (non-condensing)
Permissible storage
conditions
-20 °C to +55 °C, ≤85 % relative
air humidity, 800-1050 hPa ambi-
ent pressure
Power supply
2 x 1,5 V
AAA batteries
Battery life
For approx. 250 measurements,
depending on the blood pressure
level and/or pump pressure
Accessories
Instruction for use, 2 x 1.5 V AAA
batteries, storage box
Classification
Internal supply, continuous oper-
ation, type BF applied part, IPX0,
no AP or APG
Technical information is subject to change without notifi-
cation to allow for updates.
• This unit is in line with European Standard EN 60601-
1-2 and is subject to particular precautions with regard
to electromagnetic compatibility (EMC). Please note
that portable and mobile HF communication sys-