beautypg.com

Specifications – Monroe Electronics Electrostatic Voltmeter - Isoprobe® - model 244A User Manual

Page 5

background image

5

SECTION 1

SPECIFICATIONS

A.

APPLICATIONS:

The Model 244A instrument permits accurate voltage measurements of
electrostatic or other high impedance sources without physical contact. It utilizes
an electrostatic chopper for low drift, and negative feedback for accuracy and
probe-to-surface spacing insensitivity.

B.

FEATURES:

Integrated Solid State Circuitry

Good Accuracy

Stability
Reliability

Easy Maintenance

±3kV Input Range

Fast Speed-of-Response

Specified AC Performance

Low Noise

Digital Liquid Crystal Display

HV/1000 Recorder Output

Connectable for unlimited multiple channel operation

C.

RANGE: 0 to ±3000 volts DC.

D.

STATIC ACCURACY:

The recorder output accuracy referred to the input is better than 0.10% of reading
or 0-1V whichever is greater. (Probe-to-surface spacing .020”, 0.5mm).

E.

SPEED-OF-RESPONSE:

10-90% probe-to-surface spacing .040" (1.02mm), gain adjusted for critical
damping.

Speed of Response

typ.

max.

1kV Step .......................................... 1.5 msec ................. 2.5 msec

2kV Step .......................................... 2.0 msec ................. 3.5 msec

3kV Step .......................................... 3.0 msec ................. 4.5 msec

F.

SETTLING TIME:

typical - 3.5msec. 0-99% 1kV step, 040"(1.02mm) gain adjusted for critical

damping.

15 msec. 0-99.9% 1kV step, .040"(1.02mm)