beautypg.com

Monroe Electronics Electrostatic Voltmeter - Isoprobe® - model 244A User Manual

Page 22

background image

21

APPENDIX IV

MODELS 1017E/22D AND 1017S/22D GRADIENT

ADAPTERS

A. SPECIFICATIONS:


Range:

With Model 244 or 279; 0 to ±3000 volts/mm (Refer to Note 1)

Accuracy:

Better than 2% (Refer to Note 2)

Drift:

<0.01 V/mm per hour (Refer to Note 2)

Noise:

With Model 244 or 279; <0.3 V/mm RMS <2V/mm p-p referred to input

(Refer to Note2)


Speed of

Response:

With Model 244 or 279; <4 mS max. 3 mS typ., 10% to 90%

Resolution:

Approx. 3 cm dia. spot (Refer to Note 2)


Dimensions: 1

9

/

32

” (40mm) dia. x 1

1

/

8

” (29mm) high


Attitude:

End-view is axial

Side-view is radial

NOTES:

1.) Limited to approx. 2kV/mm air gap breakdown

2.) Specified uncontaminated atmosphere at 1 centimeter probe-to-surface

separation

B. DESCRIPTION:

These adapters are designed to convert an ISOPROBE

Electrostatic Voltmeter

into an Electrostatic Fieldmeter. This fieldmeter mode of operation makes
possible non-contacting measurement of surface potentials of virtually any
magnitude so long as the probe is mounted at the proper distance.

The adapters consist of a metal cup with a carefully chosen aperture, through
which the Model 1017, or 1034 probe “looks”, attached to a small insulated box
designed to slip over the end of the probe and a 10 foot flexible grounding wire
which is attached to the metal cup.

C. INSTALLATION:

1017E/22D: Using a 1/16” Allen wrench, retract the setscrew enough to allow the end of
the probe to slide in to the square opening until the apertures align. Gently re-tighten the
setscrew against the probe case and back off 1/4 turn. DO NOT OVERTIGHTEN. The
adapter is now prepared for calibration.