beautypg.com

ABUS Technologies Temp DM WM Temperature Transmitter User Manual

Page 9

background image

ABUS TECHNOLOGIES INC.

9

Temp DM-

WM

ATxConfig Main Window

The fields, in the above ATxConfig main window, are described as under:

1. Measuring range: Defines the beginning and the end of the transmitter humidity

and temperature range.

a. Lower Value: sets the value of the input signal associated to the 4 mA output.

b. Upper Value: sets the value of the input signal that will correspond to the 20

mA output.

When the lower limit is defined with a value higher than the higher limit the output

current operates from 20 to 4 mA. The values configured in these fields can not be

beyond the sensor measuring range. The minimum span value has to be observed as

well.

2. Sensor Failure Detection: Establishes the transmitter output behavior (upscale or

down-scale) in the presence of a sensor fail. When Minimum is selected, the output