Appendix a : specifications (continued), Communicating to a pc (continued), Appendix a : specifications – Test Products International 714 User Manual
Page 29

Appendix A : SPECIFICATIONS (Continued)
Flue Temperature Probe
Construction
Pistol Grip with Stainless Steel Shaft
Hose Length
2500mm
Insertion Length
200mm
'K' Type Thermocouple Accuracy
+/- 0.3%, +/- 1°C
Maximum Temperature
800°C
Gases
Range
Resolution
Accuracy
Oxygen (O2)
0-25%
0.1%
+/- 0.3%
Carbon Monoxide (CO)
0-10,000 ppm
1 ppm
+/- 5 ppm or 5%
Whichever is greater
Nitric Oxide (NO)
0-5,000ppm
1 ppm
+/- 5 ppm or 5%
Whichever is greater
Nitrogen Oxides (NOx)
0-5,250ppm
1 ppm
Calculated
Carbon Dioxide
0-25%
0.1%
Calculated
CO/CO2 Ratio
0-0.999
0.001
Calculated
Combustion Efficiency
0-100%
0.1%
Calculated
Pressure Measurement
Selectable Ranges
mbar, kPa and inH2O
Range
- 150 mbar to + 150 mbar
-15 kPa to + 15 kPa
-60 inH
2
O to 60 inH
2
O
Resolution
0.01 mbar
Accuracy
+/- 0.5% fsd
COMMUNICATING TO A PC (Continued)
5. To return to normal operation press and hold the Print/Comm Key
until a beep is heard and ‘COMM’ is no longer displayed.
Requests for stored and real data can be made from the PC through
Hyper Terminal to the 714.
Requesting Sored Data
1. Follow steps 1 through 3 on page 21.
2. From your computer type QTS (All Caps) and press ENTER.
3. The 714 will send stored data and return to normal operation.
Requesting Real Data
1. Follow steps 1 through 3 on page 21.
2. From your computer type QTR (All Caps) and press ENTER.
3. The 714 will send Real time data and return to normal operation.
If using the optional PC software part number A772, please refer to the
instructions that comes with the software.
Appendix A : SPECIFICATIONS
Instrument
Operating Temperature Range
14°F to +122°F (-10°C to +50°C)
Battery
Rechargeable Ni-MH
Battery Life
> 6 Hours
Charger Input Voltage
115V or 230V : 50/60 Hz AC
Fuels
Natural Gas, LPG, Light Oil, Heavy Oil
Pressure Ranges
mbar, kPa & inH2O
Display
Backlit LCD
Data Storage
10 sets of readings
Data Logging
400 sets of readings on 'Timed Interval
Logging'
Time & Date
24 Hour Real Time Clock
Dimensions
200mm x 90mm x 60mm
Weight
500g
Casing
Rubber Boot as Standard
Switch Off
Failsafe
Exhaust
Safety Spigot
Conforms to
BS7927 (and the draft BS7967)