beautypg.com

Printing data, Recalling data – Test Products International 714 User Manual

Page 25

background image

9. PRINTING DATA

WARNING: - To operate correctly there must be a clear line of sight between
the Infrared Window on the instrument (see 3.3) and the Infrared Window on
the IR Printer (see Printer instructions). Place the printer so it is no closer
than approximately 6 inches from the analyzer.

1. Press the Print/COM Key once

'rEAL' and “Stor’ will be displayed and ‘rEAL’ will be flashing.
Selecting ‘rEAL’ will cause the 714 to print the test data currently on the
display (real time data). Selecting ‘Stor’ will cause the 714 to print data
stored in memory. Use the Up and Down Arrow Keys to select the desired
print mode and press the Scroll/Enter Key.

REAL Print Mode - If rEAL print mode is selected, the analyzer will dis-
play ‘Wait’ ‘out’ and bars will change position on the LCD until printing is
complete.

After printing is complete, ‘End’ will be displayed and ‘YES’ (or ‘no’) will
blink. Selecting ‘YES’ will return the 714 to normal operation, selecting
‘no’ will return the 714 to the print selection screen so you can print
again. Use the Up and Down Arrow Keys to make the selection and press
the Scroll/Enter Key.

STOR Print Mode - If Stor mode is selected, 'Addr' will be displayed on
the top line along with 'SA ' and a location number from 0 to 9 will be
flashing on the screen. Select the required address location that you wish
to print data from by pressing the Up and Down Arrow Keys and press
the ‘Scroll/Enter’ Key. The analyzer will display ‘Wait’, ‘out’ and bars will
change position on the LCD until printing is complete.

After printing is complete, ‘End’ will be displayed and ‘YES’ (or ‘no’) will
blink. Selecting ‘YES’ will return the 714 to normal operation, selecting
‘no’ will return the 714 to the print selection screen so you can print
again. Use the Up and Down Arrow Keys to make the selection and press
the Scroll/Enter Key.

Combustion and pressure/draft printouts must be performed separately.
Perform a combustion test and save the data. Perform a pressure/draft
test and save the data in a different address. Print each address sepa-
rately.

8. RECALLING DATA

1. Press the Recall Key once

'Addr' will be displayed on the top line along with 'SA ' and a location
number from 0 to 9 will be flashing on the screen. Select the required
address location that you wish to save the data to by pressing the Up and
Down Arrow Keys

2. Press the Scroll/Enter Key once

The Time & Date of the Saved Data from the selected address location will
be displayed flashing on the screen.
The rest of the Saved Data at this address location can be reviewed by
pressing the Up and Down Arrow Keys

4. Press the Scroll/Enter Key once when you are finished reviewing the data.

'End' will be displayed with 'YES' flashing

5. Press the Scroll/Enter Key once to EXIT

or

6. Press the Up or Down Arrow Keys

'End' will be displayed with 'no' flashing

7. Press the Scroll/Enter Key once to CHOOSE another address location to

review and repeat steps 2 to 5