Page 6: Problem what to check
Serial Data
4800 BAUD
8 Data Bits
No Parity
1 Stop Bit
16 characters followed by a carriage return (0D
H
) and line feed (0A
H
)
Message length ( 18 characters) 45 MSEC
CABLE
To Starrett Collection Devices - No. PT62103 cable
Directly to PC - No. PT62103 cable and PT61768 Adapter
SERVICE INFORMATION
The height gage has been properly assembled and calibrated at the factory. There are no adjust-
ments that can be made without affecting the accuracy of this instrument, which may incur costly
repairs. If the gage has been disassembled, repair charges will be required. If a malfunction
occurs, consult the following checklist to insure that the problems are not due to improper opera-
tion of the gage.
Problem
What to Check
Does not count
Is the gage in “HOLD” mode - Check by depressing “HOLD” button
and moving slide.
Error Display
Over Speed Indicator
The word “ERR04” is displayed when the gage slide is moved at
excessive speed. This feature will cancel the display rather then
show an incorrect reading.
Loss of Accuracy
1.
Is gage base and surface plate clean?
2.
Is indicator or scriber securely fastened in place?
3.
Is the indicator or scriber clean and free of debris?
4.
Is accuracy being checked against a calibrated standard?
5.
Is the gage being used in an extremely humid environment?
If the height gage is still inoperative, return it to the manufacturer for repair. Please include a letter
of explanation and ship to:
Repair Department
The L.S. Starrett Company
165 Crescent Street
Athol, MA 01331-1915
Tel. (978) 249-3551