beautypg.com

Resetting pressure sensors, Factory settings, Resetting the external vacuum sensor – REFCO DIGIMON4 User Manual

Page 18

background image

English Instruction Manual DIGIMON-SE / DIGIMON4

16

Resetting pressure sensors

By pressing the

and

buttons the DIGIMON pressure displays are

reset to zero, in order to avoid incorrect measurement values.

To obtain a correct measurement value on the display, the DIGIMON

should not be reset when pressurised or under a vacuum.

Factory settings

Set the DIGIMON to the "Digital display" screen.

Press the

and

buttons at the same time for 5 seconds.

Display shows - - - - - - - -

Press the

button to exit.

Resetting the external vacuum sensor

Switch off the DIGIMON

Connect vacuum sensor

Hold down the

button and switch on the device using the

button.

Display shows VAC-ZERO

Press the

button to exit  display shows End.

The device can be switched on using the

button and is ready for

operation once again.






ON/OFF

SET

ON/OFF

VAC

SET

VAC

LIGHT

LIGHT

SET

This manual is related to the following products: