Mcm installation, Factory default settings – Parr Instrument Motor Control Module (MCM) User Manual
Page 7
![background image](/manuals/740426/7/background.png)
MCM Installation
w w w . p a r r i n s t . c o m
7
Factory Default Settings
MCM Module
Keys command:
1.
Press “SET” to select
2.
Press return key move to next operation mode
3.
Up/Down arrow keys to adjust value or select type
Main Screen: SP = 0
Press return key and release
Operation Mode
Select type/value Comment
r-S
Run
Run/Stop
SP
0
Decimal point position
LoC
OFF
Lock mode (lock all keys or only up/down arrow able to use)
Out1
- (read-only)
Motor output %
Press and hold down “SET” for 5-sec
Operation Mode
Select type/value Comment
InPt
v5
Input type
(v5 = 0-5V, v10 = 0-10V, nA0 = 0-20mA, nA4 = 4-20mA, nV =0-50mV)
tP-H
2206
Upper-limit range
tP-L
0
Lower-limit temperature range
CTRL
PID
Control mode (ON/OFF, MANUAL, PID and PID PROG)
S-HC
Heat
Heat/Cool control
ALA1
0
Alarm mode
SALA
OFF
System alarm
CoSH
ON
Communication write function that able to use set point from software
C-S1
RTU
Format type
C-no
2
Communication address
bPS
9600
Communication baud rate
Len
8
Data length setting
PrtY
Even
Parity bit setting
StoP
1
Stop bit setting
Press “SET” and release
Operation Mode
Select type/value Comment
At
OFF
Auto Tuning ON/OFF
PID0
0
The 0th PID Parameter
SV
0
The 0th SV
P0
170
Proportional control
C0
1
Integral control
D0
0
Derivative control
CoF0
0
Integral value
HtPd
.5
Heat/Cool cycle control
tPoF
0
Inaccuracy adjustment