Parr Instrument Primary Temp Meter User Manual
Page 4
![background image](/manuals/740416/4/background.png)
- 4 -
PTM Installation in a 4848
Factory Default Settings for Primary Temperature Module
Keys command:
1. Press “SET” to select
2. Press return key move to next operation mode
3. Up/Down arrow keys to adjust value or select type
Press return key and release
Operation
Mode
Select type/value
Comment
r-S Run
Run/Stop
SP
0**
Decimal point position
AL1H
375
Upper alarm setting
AL1L
0
Lower alarm setting
LoC
OFF
Lock mode(lock all keys or only up/down arrow
able to use)
Out1
- (read-only)
Heater output %
Out2
- (read-only)
Cooling output %
**
If using a type-J or type-K thermocouple, SV = 0
If using an RTD, SV = 1
Press and hold down “SET” for 5-sec
Operation
Mode
Select type/value Comment
InPt J**
Input
type
tPUn
C
Temperature unit
tP-H
800**
Upper-limit range
tP-L
0
Lower-limit temperature range
CTRL
PID
Control mode (ON/OFF, MANUAL, PID and PID
PROG)
S-HC
H1C2
Output 1 is Heating and output 2 is Cooling
ALA1
6
Alarm operates when PV value is higher than AL1H
setting or PV value is lower than AL1L setting
SALA OFF
System
alarm
CoSH
ON
Communication write function that able to use set
point from software
C-Sl ASCII Format
type
C-no 1
Communication
address
bPS
9600
Communication baud rate
Len
7
Data length setting
PrtY
Even
Parity bit setting
StoP
1
Stop bit setting
**
If using a type-J thermocouple, InPt = J, tP-H = 800
If using a type-K thermocouple, InPt = K, tP-H = 800
If using an RTD, InPt = Pt, tP-H = 600