Operation – LW Measurements HRB 224 Magnetic Force Balance User Manual
Page 13
Operation
External Calibration
The weight required for calibration/adjustment is defined in the factory settings (see “Specifications”).
Step Key (or instruction)
Display/Data output
1. Zero the balance
TARE
.
g
2. Start calibration
[CAL]
PLEASE WAIT
The balance will show
“PLEASE WAIT”,
after stable it will Display the
prompted calibration weight
(In this sample, 200g)
.
g CAL
4. Apply the prompted calibration
.
g
CAL
weight
5.Calibrate Weight
[CAL]
.
g
The balance will show the weighing
value and blink. After stable, the
balance will beep a sound and return
to the normal weighing mode to finish
the calibration
6. Remove the calibration weight.
.
g
11