LW Measurements LBS Large Bench Scale User Manual
Page 13

25
−
Press the HOLD key again to enter the next menu, the 
display will show CAP. Press the UNIT key to select 
full capacity. Press the UNIT key to move the flashing 
digit to the right, press the ZERO key to increase the 
flashing digit, and press the HOLD key to confirm the 
selection. 
 
−
Press the HOLD key again to enter the next menu, the 
display will show CALU=. Press the UNIT key to 
select calibration unit kg or lb (lb cannot be selected if 
lb was disabled in the Function setting – Working 
mode Sec 4.2.3). Press the HOLD key to confirm the 
selection. 
 
−
Press the HOLD key again to enter the next menu, the 
display will show CAL. Press the UNIT key to set the 
calibration weight. Press the UNIT key to move the 
flashing digit to the right, press the ZERO key to 
increase the flashing digit. Press the HOLD key to 
confirm the selection. NOTE: The calibration weight 
should be more than 2/3 of the full capacity. 
 
−
Press the HOLD key again to enter the next menu, the 
display will show DIV. Press the UNIT key to select 
division. Press the HOLD key to confirm the 
selection. 
 
−
Press the HOLD key again to enter the next menu, the 
display will show Adb. Press the UNIT key to select 
(19.20). Press the HOLD key to confirm the selection. 
 
−
 Now start the calibration following the next steps: 
 
−
Press the PRINT key, the display will show CAL and 
then the AD value. Press the UNIT key when the 
stable indicator light is displayed, it will show the 
flashing calibration weight. 
26
−
Place the known test weight on the platform, press the 
UNIT key after the stable indicator light is displayed, it 
will show ------, and then the AD value. 
−
Turn off the weighing machine, remove the test weight 
from the platform. 
−
Press ON/OFF key to turn the machine on to test if 
the weighing is accurate, if not, repeat above steps. 
 
 