Resusci Anne

5

SkillGuide
Consistent and objective real time feedback on performance
during practice helps develop and reinforce skills.
The SkillGuide provides feedback to the instructor and student
to support the student’s initial skill development by means of a
colour light display.

Inflation
When lung inflation volume reaches 0.5 l, the green signal light
will be activated. If lung inflation volume exceeds 0.8 l, an orange
warning light is activated. A red warning light indicates a too fast
inflation.

Chest compression
When the breastbone area of the chest is depressed about
50 mm (2”), the green signal light is activated.

Note: The maximum compression depth indicator is de-activated on
this product in accordance with 2010 Guidelines.

Incorrect hand position
Sensors are located under the switch cover mounted on the
underside of the chest cover. The red warning light will be
activated if the chest is depressed with hands placed outside
the correct area.

To use
See 2) under Unpacking and set up on page 4.
The SkillGuide will turn “off” automatically if no activity takes
place for 5 minutes.
Note: If the SkillGuide is turned “on” and the red signal light for
wrong hand position comes on, the SkillGuide is not completely
plugged into the manikin.

Power Supply
The four 1.5 V alkaline batteries, size C, will provide approxi-
mately 125 hours of use in the classroom. This will vary depen-
ding on use patterns and brand of batteries.

Connection Cable
This is permanently connected to the SkillGuide and has an
8 pin male connector that is to be connected to the manikin.
The connector has been moulded for proper orientation prior
to plugging it into the manikin’s side.

Caution
Make sure that the SkillGuide is unplugged from manikin:
- when airway is changed, or
- when internal service is carried out.

SkillGuide

ResusciAnne

Light

Indication

Limits*

Inflation volume:

green

correct

0.5 l - 0.8 l

orange

too much

>0.8 l

Compression depth:

green

correct

50 mm (2”)

Note: The maximum compression depth indicator is de-activated on this product in accordance
with 2010 Guidelines.

Wrong hand position:

red

Pressure outside

of acceptable area

Too fast inflation:

red

Correct ventilation <1 sec

*Measurement tolerance in the manikin: +/- 15% in normal room temperature.
Measurements based on ILCOR guidelines 2010.

Note: Maximum compression depth is approximately 55mm.