Appendix d: quick chart – Labconco XPert Nano Enclosures 38876 Series User Manual
Page 67
![background image](/manuals/738725/67/background.png)
Product Service 1-800-522-7658
62
Appendix D:
Quick Chart
Model Size
2'
3'
4'
5'
6'
Sash Height from Work Surface (inches)
9.44
9.44
9.44
12.00
12.00
Total Open Area with Bypass (sq. ft.)
1.45
2.20
2.95
4.80
5.80
Exhaust Airflow Volume at 90fpm (CFM)
130
200
265
430
520
Alarm Airflow Volume at 60 fpm (CFM)
85
130
175
290
350
Initial Mag. Gauge Pressure Reading (inches H
2
O)
0.20 to .045
0.15 to 0.40
0.10 to 0.35
0.10 to 0.50
0.10 to 0.50
Number of Laskin Nozzles needed at 10 psig
1
1
1
1
1
Exhaust ULPA Filter w/ Gasket Dimensions (inches)
18x18x3.31
30x18x3.31
42x18x3.31
(2) 24x18x3.31
(2) 18x30x3.31
*Motor Horsepower (HP)
0.13
0.13
0.30
0.60
0.60
*Capacitor (MFD, Volts)
10MFD, 115V
(2) 4MFD, 230V
10MFD, 115V
(2) 4MFD, 230V
15MFD, 115V
4MFD, 230V
15MFD, 115V
4MFD, 230V
15MFD, 115V
4MFD, 230V
Fluorescent Lights
F15T8-SP35
F17T8
F25T8
F32T8
F40T8
To determine the actual face velocity at the sash opening, airflow velocity readings
will need to be taken. This should be done across the sash opening of the
enclosure. The “average face velocity” is achieved by taking readings in two rows
across the enclosure with the readings 6" from the ends and evenly spaced every
12"; the first row is 3" down from the upper sash foil and the second row is 3" up
from the work surface.