Initial certification, Re-certification, Fluorescent light replacement – Labconco XPert Nano Enclosures 38876 Series User Manual
Page 44
Chapter 6: Maintaining the Nano Enclosure
Product Service 1-800-522-7658
39
Initial Certification
The filtered enclosure has been certified at the factory for an inflow velocity of
90±10 fpm along with the Filter Leak Test. The filtered enclosure should be
certified for the proper inflow velocity required by your Safety Officer. It is also a
conservative recommendation to perform the Filter Leak Test again should there
be any damage caused during transport.
Re-Certification
Under normal operating conditions, the enclosure should be recertified at least
annually if serviced. The certifier should perform the following tests.
Inflow Velocity Test
ULPA Filter Leak Test
In addition, the following tests should also be performed at the user’s discretion:
Measure of Line Voltage and Current
Smoke Test to determine proper airflow patterns
Lighting Intensity Test (when appropriate)
Noise Level Test (when appropriate)
Vibration Test (when appropriate)
Fluorescent Light Replacement
1. Disconnect the power.
2. Locate the small light reflector located under the control panel shown in
Figure 6-2 and Figure 6-3.
3. Remove the light reflector support by removing two Phillips screws on the
bottom of the light reflector.
4. Rotate and remove the old fluorescent lamp.
5. Reinstall the new fluorescent lamp and light reflector in reverse order.
6. Power the unit up and try the new fluorescent lamp.