Validating the glove box – Labconco Precise Controlled Atmosphere and Basic Glove Boxes 5220131 User Manual
Page 24
Chapter 3: Getting Started
Product Service 1-800-522-7658
20
Figure 3-10
Typical Tubing Connections for Dual Moisture/Solvent
Removal System Shown with the Automatic Pressure Controller
(Tubing Connection Kit 5242500 provides full details)
Validating the Glove Box
Each Precise Glove Box has been helium leak tested at the factory for
integrity and found to have no leaks greater than 1 x 10
-4
cc/sec while at a
positive pressure of 5 inches helium. Refer to Chapter 6, Initial Certification
and Appendix G for Helium Leak Test Method.
Overall Glove Box Main Chamber Leak Test
The Precise Controlled Atmosphere Glove Box has been tested with an oxygen
monitor detecting oxygen levels as low as 1 ppm; the rate of increase in
oxygen over time was determined in accordance with the ISO 10648-2 oxygen
leak decay test method. The results validate that the Precise Controlled
Atmosphere Glove Box meets Class 1 conditions for controlled atmosphere
- Precise Controlled Atmosphere and Basic Glove Boxes 5220130 Precise Controlled Atmosphere and Basic Glove Boxes 5220121 Precise Controlled Atmosphere and Basic Glove Boxes 5220120 Precise Controlled Atmosphere and Basic Glove Boxes 5220100 Precise Controlled Atmosphere and Basic Glove Boxes 5220031 Precise Controlled Atmosphere and Basic Glove Boxes 5220030 Precise Controlled Atmosphere and Basic Glove Boxes 5220021 Precise Controlled Atmosphere and Basic Glove Boxes 5220020 Precise Controlled Atmosphere and Basic Glove Boxes 5220000