Triad freeze dry system components, Setting up your triad freeze dry system, Component orientation and hose connections – Labconco FreeZone Triad Freeze Dry System 74000 Series User Manual
Page 12

Chapter 3: Getting Started
Product Service 816-333-8811 or 1-800-522-7658
8
Triad Freeze Dry System Components
Locate the model of freeze dry system you received in the following table.
Verify that the components listed are present and undamaged.
If you did not receive one or more of the components listed, contact
Labconco Corporation immediately for further instructions.
Power Cords
Catalog #
Description
Volts Hz
13
36
50
0
13
42
10
0
13
32
60
1
13
32
70
1
7400030
Triad Freeze Dryer System
230
50
X
7400040
Triad Freeze Dryer System
230
60
X
7400060
Triad Freeze Dryer System
230
50
X
7400070
Triad Freeze Dryer System
230
50
X
Plus the following:
Part #
Qty.
Component Description
7373300
1
User’s Manual
7373436
1
Tubing
1488800
2
Clamp
1291000
1
Female IEC Plug
Setting Up Your Triad Freeze Dry System
Component Orientation and Hose Connections
A minimum of 3" should be allowed between the back of the freeze dry
system and the adjacent wall surface and between the sides of the freeze
dryer and the adjacent wall surfaces. Restriction of the airflow through the
cabinet during operation could adversely affect performance.
Position the vacuum pump near the vacuum exhaust port on the rear of the
cabinet. Attach the supplied hose to the exhaust port using the clamp
provided. Attach the other end of the hose to the inlet fitting on the vacuum
pump. Secure with the clamp provided. The hose may be cut to proper length
to allow for gentle bends without kinks.
Connect the power cord from the vacuum pump into the outlet on the back of
the unit. If your vacuum pump power cord does not have a female IEC plug,
replace the cord end with the female IEC plug provided with the unit (Part
number 1291000).