Labconco FreeZone 2.5 Liter Freeze Dry Systems 76705 Series User Manual
Page 39

Chapter 7: Troubleshooting
Product Service 816-333-8811 or 1-800-522-7658
35
III. System Components and Collection Chamber
Isolation
This test determines if the source of a leak is in the drying chamber.
1.
Remove the drying chamber or manifold.
2.
Leave the gasket on and turn a large freeze dry flask upside down to cover
the connection port.
3.
Start the Freeze Dryer and observe the vacuum indication. If the vacuum
indication is good, then the problem is in the drying chamber or manifold
and you should return to Section II, Gaskets, Tubing, Connections, Sample
Valves.
Freeze Dry Flask
Isolating the freeze dry system from the accessory chamber/manifold
Gasket
This manual is related to the following products: