Chapter V
Maintenance and Troubleshooting Information
Troubleshooting the Instrument 
 
The advanced design of the RMLD-IS makes it one of the most reliable leak survey 
instruments available. However, should you experience problems with the instrument or 
suspect that the instrument is not operating properly, do not use the instrument for leak 
survey work until the problem is resolved. 
 
Only a qualified RMLD-IS repair technician should attempt to repair or adjust the 
instrument. There are no user serviceable components in the RMLD-IS that can be 
repaired or replaced. 
 
Most often, a problem has a simple cause. The following table provides a list of common 
problems, cause and solution. Should you have a problem not listed or the solution 
doesn’t work, please contact HEATH for further assistance. 
 
Symptom
Probable Cause(s)
Solution
Readings are higher than 
normal at short range and 
lower than normal for long 
range. 
 Laser calibration has
drifted
• Run Self-Test
Concentration reading low 
and will not pass the Self-
Test 
 Laser calibration has
drifted
• Run Self-Test and use
the override function
Have to use Self-Test 
override each time the 
Self-Test is performed 
 Laser calibration has
drifted
 
 Laser calibration drift 
rate is too high
• Perform the Self-Test
on a daily basis
 
• Contact HEATH. 
Unit will not turn on
 Low battery
• Recharge internal
battery, or
 
• Use the External Power 
Supply
Keep getting a warning 
sound or icon when 
scanning 
• Scanning at a distance
beyond the RMLD-IS 
range 
 
• Background surface is 
absorbing or reflection 
the light level 
 
• Low battery 
• Move closer to the
target, or
• Change angle to the
target to get a better 
reflecting background 
• Check battery level and
recharge if necessary